Establishing metrology standards in microfluidic devices: MFMET II
Author
Celeste Chidiac, PhD
Publication Date
June 26, 2025
Keywords
Metrology
Standardization
Flow measurement
Measurement protocols
Validation methods
Reliability
Reproducibility
ISO standards
Transfer standards
Your microfluidic SME partner for Horizon Europe
We take care of microfluidic engineering, work on valorization and optimize the proposal with you
Turning microfluidics into a standardized, trustworthy technology
Microfluidics has progressed quickly thanks to advances in microfabrication and nanotechnology. But while devices are increasingly available, reliable qualification and characterization methods remain lacking at the system level.
EURAMET 24NRM03 MFMET II is a European project focused on establishing metrology standards in microfluidic devices, including complex systems used in Organ-on-Chip (OoC), diagnostics, microreactors, and other miniaturized lab platforms.
MFMET II addresses this gap by developing protocols, guidelines, and metrology tools that help the field move from “each product has its own validation method” to shared, standardized, comparable testing.
The project focuses on turning key measurement needs into practical protocols and guidance that can be used consistently across labs, manufacturers, and application domains. A key part of that ambition is feeding directly into the work of standardization bodies like ISO/TC 48/WG 3, helping translate metrological rigor into practical industry guidance.
MFMET II project in a nutshell
MFMET II aims to strengthen the metrological foundations needed for reliable standardization in microfluidics and organ-on-chip technologies.
A central objective is to define standardized measurement procedures for core microfluidic quantities that drive performance and comparability, including particle-laden flows (droplets, bubbles, particles, cells), shear stress, pressure drop, flow resistivity, and device volumes (dead and total volume). This helps move the field away from “device-specific” validation toward methods that are repeatable and comparable across platforms.
In parallel, the project will develop protocols for integrating sensors, actuators, and fluidic components using scalable and sustainable manufacturing approaches, while addressing sterilization, contamination prevention, and material-medium interactions, such as absorption/adsorption, coatings, and wettability changes, which are especially critical for organ-on-chip systems.
Finally, MFMET II will establish guidelines for quality control and reliability (including test approaches, failure modes, and safety aspects). The project will also develop an integrated microfluidic setup with multiple sensors and actuators intended to act as a metrological transfer standard. The project’s outcomes will be disseminated through publications, protocols, white papers, and recommendations aimed at regulators, standardization bodies, and the global microfluidic community.
How does MIC contribute to MFMET II?
MIC’s role in MFMET II is to bring hands-on microfluidics and flow-control expertise to the project, ensuring that emerging metrology protocols translate into practical, system-level qualification methods that work in real devices and real workflows. MIC helps stress-test and refines procedures around stable low-flow operation, pressure-flow characterization, and repeatable flow verification. Crucially, this includes accounting for the failure modes that quietly undermine reproducibility in practice: clogging, drift, and subtle changes in the fluidic path that can be easy to miss but hard to explain away.
A good illustration is MIC’s microfluidic flow sensor, designed for wide-range, high-accuracy liquid flow measurement in microfluidic setups and explicitly aimed at making flow control and verification more trustworthy across experiments and platforms.
Download the MIC Horizon Europe 2026/2027 Calls Calendar:
Funding and Support
The project has received funding from the European Partnership on Metrology, co-financed from the European Union’s Horizon Europe Research and Innovation Programme and by the Participating States.
Start date: June 2025
End date: June 2028
About the author
Céleste Chidiac, PhD
Scientific writer and digital marketing specialist
Céleste Chidiac holds a doctorate in neuroscience and molecular biology from the University of Strasbourg. Since joining the Microfluidics Innovation Center in 2023, she has combined scientific writing with website management. Her reviews draw on published research to explain microfluidic technologies, compare experimental approaches and discuss their applications and limitations. She helps researchers understand how these methods work and assess their relevance to their own research.
Check our Projects
FAQ – MFMET II and microfluidic measurement quality
Practical distinctions for interpreting measurements, comparing devices and understanding how metrology supports microfluidic standardisation.
How do calibration and validation differ?
Calibration links instrument readings to reference values with associated uncertainties. Validation establishes whether specified requirements are met and are suitable for the intended application.
Can a repeatable measurement still be wrong?
Yes. Closely grouped readings can share a systematic bias. Repeatability describes consistency under specified conditions; it does not, by itself, establish accuracy.
What makes a measurement result traceable?
Traceability requires a documented, unbroken calibration chain linking the result to a specified reference, with uncertainty evaluated along that chain. It belongs to the result, not merely the instrument.
Why is sensor resolution not the measurement uncertainty?
Resolution concerns the smallest change in the measured quantity that produces a perceptible response. Measurement uncertainty also includes relevant contributions from calibration, repeatability, environmental conditions and the measurement method.
Why assess the complete fluidic assembly?
Connections, tubing and operating conditions can change performance after components are combined. Testing an assembled system checks effects that individual sensor calibrations cannot establish.
Is a flow-rate value enough to specify shear stress?
No. Wall shear stress also depends on fluid viscosity and channel geometry. Identical flow rates can therefore produce different mechanical conditions in different devices.
Why does internal volume matter when switching liquids?
Liquid already inside channels, connectors and chambers delays replacement. Poorly flushed regions can prolong carryover, so changing the inlet liquid does not ensure immediate exchange downstream.
How can chip materials affect the delivered sample?
Molecules may adsorb onto surfaces or be absorbed into materials, changing sample composition. Compatibility testing must consider the actual liquid, surface treatment and exposure conditions.
What does a microfluidic transfer standard do?
A characterised transfer standard provides a common basis for comparing measurement systems. Its stability, operating conditions and associated uncertainties must be established for those comparisons.
Do MFMET II protocols automatically become ISO standards?
No. Project protocols can inform standardisation, but ISO standards require a separate development and consensus process. A project report is not automatically an adopted standard.